+91-9116607583 |
sales@tempsens.com |
Specifications
Type | Portable |
Certification | CE Certified |
Application | Temperature Reading |
Weight | 15 Kg |
Display Type | Digital |
Frequency | 50Hz |
Driven Type | Electric |
Temperature Range | -196°C To -80°C |
Brand Name | Tempsens |
Temperature Resolution | 1°C |
Control Sensor | RTD Pt100 |
Method of Control | Digital PID Controller |
Computer Interface | RS 232 |
Download | ![]() |
Additional Information
Min. Order Quantity | 1 Piece(s) |
Port | Mumbai |
Payment Terms | T/T |
Delivery Time | 6-8 Weeks |
CALsys -196/-80 offers ultra low temperature range from -196 to -80°C. It is a highly stable standard source for calibrating Thermocouples / RTD. It has been designed for low temperature range calibration and find application in the glass, pharma, electrical power, automotive & material processing industries. The comparison volume is a metallic block of aluminium having dimension 25mm diameter with 300 mm long which fixed into the liquid nitrogen container for calibration of RTD & thermocouple. The temperature of the calibrator is set and controlled by a self tuned PI D controller with automatic super fine adjustment.
Key Features
Standard Accessories
Optional Accessories
Specification
Temperature Range | -196 to -80°C |
Temperature Resolution | 1.0°C |
Stability | ±0.1°C at-196°C ±0.1°C at -130°C ±0.1°C at -80 |
Uniformity | ±0.2°C at-196°C ±0.15°C at -130°C ±0.17°C at -80 |
Controlling Sensor | RTD PT100 |
Method of Control | Digital self tuned PID Controller |
Insert Construction | Dia 25 x 300 mm long (2 x 6 mm & 2 x 8 mm holes) of 300 mm insertion depth |
Time to Reach Max Temp | 30 Mins |
Computer Interface | RS - 232 |
Operating Temperature | 20 to 45°C |
Power Requirement | 230 VAC, 300 W |
Dimensions of Container | 580(H) x 280(Dia) |
Nitrogen Container | 10Ltr. |
Dimension of Control Panel | 270(H)x380(W)x270(D ) |
Weight | 15Kg (without packing) |