Tempsens Instruments (I) Pvt. Ltd.

Calibration Equipment (Master Thermocouple - TTC)

Calibration Equipment (Master Thermocouple - TTC)

Specifications

Certification CE Certified
Application Testing Material
Warranty Guarantee For Workmanship
Temperature Element Cr/Al K Type
Accuracy Class 1
Outer Diameter 6mm
Temperature Range 0 to 1200°C

Additional Information

Product Code TTCK
Min. Order Quantity 1 Piece(s)
Port Mumbai
Payment Terms T/T
Delivery Time 4-6 Weeks
Packaging Details Suitable Rugged Carrying case will be supplied

K type and N type Thermocouple is mainly use in Industries as a secondary master sensor. It works together with readout device to measure Temperature or change of temperature. It has wide applications for dry-wells or temperature baths. It is recommended to calibrate this Thermocouple annually over the full temperature range.

Technical Specifications

Make Tempsens

No. of Element Simplex

Temperature Range 0 to 1200 °C

Sheath Material Inconel 600

Sheath length 400 mm

Extension Cable 1.5 mtr. Long Teflon insulated cable with male/female miniature connector

Sheath Dia 6 mm

Handle Dimension 15 mm (OD) X 100 mm( L)

Calibraat 5 points at Tempsens NABL Accredited Lab

Accuracy Special Class (1.1°C or 0.4% of temperature whichever is greater)


Yes! I am interested


Description

Model No. Prefix T/CType Material Range (°C) Accuracy Sensitivity( v/°C) Options in Sheathlength (mm) Options in Cablelength(mm) Outer Sheath
TTCN/TTCK N/K NiCrSiNiSi/NiCr-NiAl Upto 1200 ±1.1°C or ± 0.4% 39 - 41 300, 450 &600 1000 & 1500 INCONEL
TTCR/TTCS R/S PtPt-10RhPt-13RhPt/ Upto 1500 ±0.6°C or ±0.1% 43628 300, 450 &600 1000 & 1500 CERAMIC
  Selection : Part No. Model No. type - Length Ex. : TTCN-300





Looking for "Calibration Equipment (Master Thermocouple - TTC)" ?