Liquid Bath Calibrator

Leading Manufacturers, Exporters, Wholesaler, Retailer of CALsys -35/200 High Stability Bath, CALsys -40/50 High Stability Bath and CALsys-80/50 High Stability Bath from Udaipur.

MOQ : 1 Piece(s)

Business Type Manufacturer, Exporter, Supplier, Retailer, Wholesaler
Temperature Range -35 to 200°C
Stability ±0.01°C at -35°C, ±0.05°C at-0°C, ±0.04°C at 200°C
Uniformity ±0.03°C at -35°C, ±0.06°C at 0°C, ±0.07°C at 200°C
Resolution 0.01°C
Cooling Time 60 min (room temperature to -35°C )
Volume 6 Ltr
Access Opening 105 (L) x 105 (W) mm
Immersion Depth 150 mm
Method of control Digital Self Tuned PID Controller
Computer Interface RS - 232
Power Requirement 230 VAC, 2 KW (50 Hz)
Dimensions 540(H) x 220(W) x 620(D)mm
Weight 65 Kg
Calibration Accredited calibration certificate provided
Environmental operating conditions 0°C to 40°C, 0% to 90% RH (non-condensing)
Specifications valid in environmental conditions 13°C … 25°C
Input (CaLsys -35/200 Autocal) Four Channels (One master and three test sensors), high-quality LEMO connector suitable for both T/C
Download View
Payment Terms T/T
Delivery Time 6-8 Weeks

Click to view more

Preferred Buyer From

Location Worldwide

CALsys -35/200 is a Highly Stable Stirred Liquid Bath for Calibration RTD / Thermocouple and also for compensation of cold junction.

CALsys -35/200 offers the temperature range of -35 to 200°C. It has large tank which is filled with liquid (Water+Methanol/Silicon Oil). The liquid is heated or cooled to the desired temperature, where Methanol+Water are used for Negative Temperature and Silicon Oil is for Heating purpose.

It has better stability & uniformity than dry block bath and as it uses of stirred liquid. This CALsys -35/200 is built using CFC-free refrigerants.

Key Features:

  • High Accuracy
  • High Stability and Uniformity
  • Wide Temperature Range
  • PC Interfacing
  • Simple to Use and Cost-Effective


Accessories
Standard Accessories

  • Reference Standard PRT…Part No.TPRT-A-300
  • NABL accredited calibration certificate - 3 point
  • Software - Including for setting bath temperature and monitoring the PV. Graphical representations of PV/TIME.
  • Operational Manual


Optional Accessories

  • Methanol............10 Ltr. No.(3050-M)
  • Silicon Oil............10L Part No.C200S-LB SO

 

Input (CaLsys -35/200 Autocal): Four Channels (One master and three test sensors), high-quality LEMO connector suitable for both T/C and Rtd.

Software (CaLsys -35/200 Autocal): The calibrator will be provided with software for data recording (Manual Mode) and report generation in Auto Mode.

Data Logging (CaLsys -35/200 Autocal): Data Logging facility with logged data export to the computer through  LAN port (Optional USB)

Read More...


Yes! I am interested


MOQ : 1 Piece(s)

Business Type Manufacturer, Exporter, Supplier, Retailer, Wholesaler
Standard CE Certified
Temperature Range -40 to 50°C
Stability ±0.04 at -40°C, ±0.07 at 0°C, ±0.05 at 50°C
Uniformity ±0.05 at -40°C, ±0.09 at 0°C, ±0.07 at 50°C In stock Part Code: CALsys -40/50 ₹150,000.00
Accuracy ±0.5 °C
Temperature Resolution 0.1°C
Cooling Time 60 Min ( room temperature to -40°C )
Medium Methanol
Controlling Sensor PT-100
Volume 9 Ltr
Access Opening 105 x 105 mm
Depth 150 mm
Method of control Digital Self Tuned PID Controller
Computer Interface RS - 232
Operating Temperature 20 to 25°C
Power Requirement 230 VAC, 2.0 KW
Dimensions 410(W) x 702(H) x 603(D) mm
Instrument Weight 41 Kg(without packing)
Download View
Payment Terms T/T
Delivery Time 6-8 Weeks

Click to view more

Preferred Buyer From

Location Worldwide

CALsys -40/50 is a Highly Stable Stirred Liquid Bath for Calibration RTD / Thermocouple and also for compensation of cold junction.

CALsys -40/50 offers the temperature range of -40 to 50°C. It has large tank which is filled with liquid Methanol. The liquid is heated or cooled to the desired temperature. It has better stability & uniformity than dry block bath and as it uses of stirred liquid.

This CALsys -40/50 is built using CFC-free refrigerants.

Key Features:

  • Large Immersion Depths
  • High Accuracy
  • High Stability and Uniformity
  • Wide Temperature Range
  • PC Interface
  • Simple to use and Cost-effective


Accessories:

  • Reference Standard PRT...Part No. TPRT-A-300 NABL accredited calibration certificate - 3 point.
  • Software - Cal Soft including for setting bath temperature and monitoring the PV. Graphical representations of PV/TIME with 2 hours data logging.
  • Operational Manual.
Read More...


Yes! I am interested


MOQ : 1 Piece(s)

Business Type Manufacturer, Exporter, Supplier, Retailer, Wholesaler
Temperature Range -80 to 50°C
Stability ±0.04°C at -80°C, ±0.07°C at 0°C, ±0.05°C at 50°C
Uniformity ±0.05°C at -80°C, ±0.09°C at 0°C, ±0.07°C at 50°C
Temperature Resolution 0.1°C
Time to Reach Min. Temp 2 Hrs
Medium Methanol
Controlling Sensor PT100
Volume 9 Ltr.
Access Opening 105 x 105 mm
Depth 200 mm
Method of control Digital self tuned PID Controller
Emissivity 0.95 ±0.01
Computer Interface RS - 232
Operating Temperature 20 to 25°C
Power Requirement 230 VAC, 2.0 KW
Dimensions 630 (W) x 1200(H) x 500 (D)mm
Instrument Weight 100 Kg
Download View
Payment Terms T/T
Delivery Time 6-8 Weeks

Click to view more

Preferred Buyer From

Location Worldwide

CALsys -80/50 is a Highly Stable Stirred Liquid Bath for Calibration RTD / Thermocouple and also for compensation of cold junction. CALsys -80/50 offers the temperature range of -80 to 50°C. It has large tank which is filled with liquid Methanol. The liquid is heated or cooled to the desired temperature. It has better stability & uniformity than dry block bath and as it uses of stirred liquid. This CALsys -80/50 is built using CFC free refrigerants.

Key Features:

  • Large Immersion Depths
  • High Accuracy
  • High Stability and Uniformity
  • Wide Temperature Range
  • PC Interface
  • Simple to use and Cost effective


Accessories:

  • Reference Standard PRT...Part No. TPRT-A-300
  • NABL accredited calibration certificate - 3 point.
  • Software - Cal Soft including for setting bath temperature and monitoring the PV. Graphical representations of PV/TIME with 2 hours data logging.
  • Operational Manual.
Read More...